6-15
IM 701610-01E
Triggering
6
Setting the Level
10. Pressing the Level/Coupling soft key displays a menu used to set the level,
coupling, hysteresis, and HF rejection.
The Level setting for each
channel can be input directly
using a USB keyboard. ( )
11. Turn the jog shuttle to move the cursor to the channel on which the level is to be
set.
12. Press SELECT to display the level setting menu.
13. Turn the jog shuttle to set the level.
You can use the arrow keys to move between the digits. In addition, pressing
RESET sets the trigger level to 0 V.
Setting the Hysteresis
14. Turn the jog shuttle to move the cursor to the channel on which hysteresis is to be
set.
15. Press SELECT to select or .
Setting the Trigger Coupling
16. Turn the jog shuttle to move the cursor to the channel on which the coupling is to
be set.
17. Press SELECT to select DC or AC.
Setting the HF Rejection
18. Turn the jog shuttle to move the cursor to the channel on which the HF rejection
(HF Rej) is to be set.
19. Press SELECT to select ON or OFF.
Setting the Number of Times Condition B is to be Met
20. If the jog shuttle control is not set to Count, press the Count soft key.
21. Turn the jog shuttle to set the count.
You can use the arrow keys to move between the digits. Pressing RESET resets
the value to 1.
Setting the Hold Off
22. Set the hold off time according to the procedures given in 6.4, “Setting the Hold Off
Time.”
Note
• The status setting of conditions A and B applies to A -> B(N) trigger and A Delay B trigger.
• The trigger level, hysteresis, trigger coupling, and HF rejection settings apply to simple trigger
and enhanced trigger.
6.8 Setting the A→B(N) Trigger (ENHANCED)