6.1.7.1.2. Fine Tuning
The purpose of this section is to adjust more carefully the "shoulder" ofthe step response.
1.
Setup the system by performing steps 1,4, and 6 from section
6.1.7.1.1.
above.
2.
Tum on the system power.
3.
Adjust the input signal amplitude and offset so that the scanner now steps from -2.0° to 0.0° in
a square wave fashion and at the same frequency as before.
4.
Adjust the oscilloscope channel monitoring Vp to lOmV/div. The system should now look
like Figure 8.
Note: Since only the coarse tuning has been performed thus far, your waveform may look
slightiy different at this point. However, the same tuning mles still apply.
5.
Tum R28 CCW to bring the first overshoot at the same level as the settled waveform. See
figure 9.
CH2
CHI
CH Imposition Out
CH2=Position In
CH2
Chi
I6mi^
k MB iiftnii/
^. M
I5ms
ixUU/
1.58
V
Chi ' lamO C MM SftftmV' ^
M
iS'ms iM\6/ 1.58
V
CH1=PositionOut
CH2=Posaion In
CH1
Figure 8. Figure 9.
6. Tum R25
CW
to eliminate the first undershoot. See Figure 10.
37