2000 MS Method Chemical
Ionization
Introduction
Chemical ionization (CI) provides mass spectral data that complement electron
ionization (EI) data for the analysis of complex compounds. In the CI mode of
operation, a CI reagent gas is introduced into the ion trap analyzer. The reagent
gas is ionized by EI. Then, the sample molecules are ionized by ion molecule
reactions with the reagent gas ions.
CI is a softer ionization technique than EI. That is, CI imparts less energy to the
sample molecules than does EI. Thus, the ionized sample molecule undergoes
less fragmentation, and an ion indicative of the molecular weight is more likely to
be observed. In addition to molecular weight confirmation, CI mass spectra often
provide other significant structural information that may not be available from EI
mass spectra.
The Chemical Ionization Scan Function
EI space charge control uses the integrated ion current from a fixed ion time
prescan to calculate the analytical scan ionization time. In contrast, CI space
charge control does not use a fixed ionization time prescan to measure the
charge formed. CI uses a portion of the ions detected during the previous
analytical scan to calculate the ionization and reaction times for the next
analytical scan.