You can use a spit ratio of 100 after injection to vent the injector of any residual
solvent.
Use a very low split ratio after flushing to conserve carrier gas.
The first row of the spreadsheet contains the Initial Split State and Split Ratio for
the 1079 injector. The 1079 will equilibrate to this setting when the method is
activated, and will be restored to this setting when each chromatographic run is
completed.
Rows after the first row of the spreadsheet contain the time-programmed settings
for the 1079 Split Ratio. The Split Valve will be switched to the ratio in each row
at the indicated time.
1177 Injector
If you indicate that a 1177 Injector is installed, an Oven Power switch, a
Temperature setting and a Split Ratio button will appear.
Indicate whether or not you will use the
installed 1177 injector by clicking on
the “On” or “Off” radio buttons.
Set the desired
Temperature.
If the 1177 Injector is connected to a Type 1
EFC, press the “Split Ratio” button to build the
Split Ratio Time Program.
NOTE: You must have already configured the Type 1 EFC in the corresponding
position in the Flow/Pressure Section before you can program the split ratio.
Specifically, to build a 1177 Split Ratio Program, a Front 1177 must have a
corresponding Front Type 1 EFC, a Middle 1177 must have a corresponding
Middle Type 1 EFC, and a Rear 1177 must have a corresponding Rear Type 1
EFC.
Use the Split Ratio spreadsheet to build a split ratio time program to control the
1177 injector / Type 1 EFC combination.