Axis BTA 2100D Manual do Operador Página 586

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 597
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 585
584
Item Description
Attenuation 1 to 4096
Sets the program attenuation.. Attenuation is relative to the
maximum full scale value of the detector. An attenuation of 1 scales
the chromatogram so that a signal 1/4096
th
of the detector's full
scale is the largest signal that will be printed.
Add Adds the active parameters to the program list.
To add a change in chart speed or attenuation to the report time
program.
Delete Deletes the highlighted line in the program.
To delete a line in the report time program, canceling that change in
chart speed or attenuation.
Results Options Dialog Box
Item Description
Units
Any units, up to 10 characters
Assigns a name for the units used in the results report.
Number of
Decimal Digits
0 to 6
Sets the number of digits in which results are to be
expressed.
Run Log If checked, includes the method and the Time Events Log at
the end of the report. Entries in the Time Events Log
correspond to the actual times when the events occurred.
Error Log Includes error messages that occurred during the run at the
end of the report. Error messages generated by post-run
Data Handling are always printed whether the error log option
is active or not.
Calibration
Report
Includes the text of the calibration block report as part of the
results report.
Vista de página 585