Axis BTA 2100D Manual do Operador Página 174

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 597
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 173
172
Segment Setpoints
Item Description
Scan Time
The duration of each scan (in seconds). Values from 0.10
to 5.00 seconds per scan. Below is displayed the number
of micro scans per scan for the selected mass range and
scan time.
Multiplier Offset
Multiplier offset (relative to 10
5
gain). Range 300 Volts.
Emission Current
Emission current may be set from 5 to 100 µA.
Count Threshold
The minimum number of intensity counts per sampling
interval required before a signal is recorded. Values from 0
to 1000 counts; it is typically set to 0 or 1 count.
Mass Defect
Mass defect is defined as the difference between the
nominal mass of an atom (or ion) and its exact mass.
Values from -300 mu per 100u to +300 mu per 100u can
be set.
Cal Gas
Click on box if Calibration gas is used in the Method
segment.
Vista de página 173
1 2 ... 169 170 171 172 173 174 175 176 177 178 179 ... 596 597

Comentários a estes Manuais

Sem comentários