Shutdown Mode
Item Description
Shutdown button
Shuts down the system in an orderly manner. The pump
speed is slowly reduced and the heaters are turned off so
that the system slowly cools.
Reset button
Brings the system back to a ready state if the Shutdown
procedure was started. This restarts the pumps and turns
on the heaters.
Diffusion Pump Shutdown Sequence
NOTE: The Diffusion Pump System shutdown sequence takes at least 30
minutes:
The Saturn GC/MS heaters are turned OFF.
The GC temperature zones should be cooled downed before maintenance is
performed.
The diffusion pump heater is turned OFF while the Peltier Baffle cooler is
kept on. The DIFFUSION PUMP STATUS changes from NORMAL to
COOLING. After 15 minutes, when there is no pump vapor to contaminate
the system, the Peltier Baffle cooler is turned OFF. After another 15 minutes,
the cooler has warmed to the point that it won’t condense water vapor, and
the pump is considered to be off. The DIFFUSION PUMP STATUS then
changes from COOLING to OFF.
NOTE: If the word “FAULT!” is displayed in the Vacuum Status box, click
Diagnostics for detailed information.
The Vacuum Status represents the state of the diffusion pump. To precisely
identify the problem, the USER should go to the Diagnostics mode and run
"System Test to Completion”. Fault messages may appear on startup or during
normal operation.