NOTE: If the excitation amplitude used is too large, the parent ion and product
ion spectra will be absent because both ions will be ejected from the trap. If the
value is too small, the parent ion spectrum will be dominant and the product ion
spectrum will be weak or missing.
The MS “q” Calculator
For convenience, The MS q Calculator is available in all Ion Preparation Options.
Storage rf values are often reported in terms of the Mathieu “q” parameter in the
scientific literature. The value of the “q” parameter determines the stability of the
parent ion trajectory. More stable trajectories allow higher excitation voltages to
be applied before ions are ejected from the trap. It has been empirically observed
that a “q” value of 0.4 provides an optimum yield of product ions. This calculator
utility determines the corresponding CID storage rf value (m/z) following user
entry of an ion mass (m/z) and desired “q”.
NOTE: Right click of the mouse on the “q” calculator button provides a “q” to “
calculator.
Customize Non-resonant Method
The parameters in the MS/MS Ion Prep Method Editor window are the basic
ones needed to create a MS/MS Ion Prep Method file. For the vast majority of
applications, no other information is required. However, some applications may
require further customization.
NOTE: Be sure that you have specified the parent ion mass before opening the
Customize Parameters window.