Axis BTA 2100D Manual do Operador Página 179

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 597
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 178
177
Ion Storage (SIS) Parameters
Item Description
Storage Mass
Ranges
Table to enter ion(s) or range of ions for selective storage.
Enter up to five mass ranges.
Low Mass
Low m/z integer mass. Mass range from 25 m/z to 650 m/z.
High Mass
High m/z integer mass. Mass range from 25 m/z to 650 m/z.
Ejection Masses
Table to enter ion(s) for selective ejection. Enter up to five
ions.
Ion Mass
Integer mass(es) of those ion(s) to be ejected. Use ADD
button to enter value.
Amplitude
Range is 1% to 200% with a default of 100%.
Customize SIS
Item Description
Ionization Storage
Level
Minimum mass that can be stored and isolated. Range: 20
m/z mass 60 m/z. Default = 48 m/z.
Autoscale
SIS waveform amplitude is automatically scaled as a function
of the frequency components and storage rf voltage selected.
By unchecking Autoscale, the user can enter a waveform
amplitude.
Vista de página 178
1 2 ... 174 175 176 177 178 179 180 181 182 183 184 ... 596 597

Comentários a estes Manuais

Sem comentários