
447
Calibration Setup
Item Description
Calibration Type % (No Calibration), Internal Standard (IS), External
Standard (ES), Normalized % (N%)
Sets the calculation type for the Method. Determines
what calculations are performed for calibrations and
analyses.
Number of Calibration
Levels
1 to 10 levels
Sets the number of calibration levels you will be using.
Default Curve Origin Include, Ignore, Force
Sets the default treatment of the origin in each curve.
Choose to include, ignore, or force the curves through
the origin.
Default Curve Fit Linear, Quadratic, Cubic
Sets the default for how each calibration curve is fit.
Choose a linear, quadratic, or cubic equation for the
curves.
View Curves... Opens the Calibration Curve window. Used to view,
edit, print, or export the calibration curve.
Comentários a estes Manuais