355
You can use a split ratio of 100 after injection to vent the injector of any residual
solvent.
Use a very low split ratio after flushing to conserve carrier gas.
The first row of the spreadsheet contains the Initial Split State and Split Ratio for
the 1177 injector. The 1177 will equilibrate to this setting when the method is
activated, and will be restored to this setting when each chromatographic run is
completed.
Rows after the first row of the spreadsheet contain the time-programmed settings
for the 1177 Split Ratio. The Split Valve will be switched to the ratio in each row
at the indicated time.
Column Oven Section
The Column Oven window contains the Column Oven Program. Use the Column
Oven Program to specify the Stabilization Time of the Column Oven, and to
specify its programmable temperature ramp.