173
Ionization Mode - EI Auto
Item Description
Mass Range
Segment Breaks
The EI scan is made up of between 1 and 6 scan
segments, each holding a contiguous mass-range
segment. Overall, the set covers the entire mass range of
10 to 650 m/z. Only the intermediate segment boundaries
can be changed by the user.
Ionization Storage
Level
The rf voltage used during the ionization period. Values
from 5 to 150 m/z.
Ionization Time
Factor
The Ionization Time Factor, given as a percent, is a
number that is multiplied by the calculated ionization time
(determined by the AGC pre-scan pulse) to give the actual
ionization time. Values from 1 to 999%.
Target TIC Value
Target TIC value determines how many ions AGC allows
into the ion trap during the calculated ionization time.
Range of values is 10 to 65000.
Maximum
Ionization Time
Sets the limit for the longest ionization time which AGC can
use. Values from 10 to 65000 sec.
Prescan
Ionization Time
This ionization time determines the number of ions formed
for the prescan. This value is then used to calculate the
ionization time for the analytical scan. Range 10 to 2500
usec.
Background Mass
The lowest mass used to determine the target TIC (total
ion current) value; all ions of lower mass are ejected from
the ion trap before the AGC prescan pulse. The
background mass can be set to any value from 10 m/z to
300 m/z; it is typically set to 45 m/z.
rf Dump Value
This value specifies the high m/z used for the prescan. It
should always be as high or higher than the analytical scan
range high mass.