Axis BTA 2100D Manual do Operador Página 185

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 597
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 184
183
Item Description
High Edge Amplitude Amplitude of broadband waveform used to eject
masses above the isolated parent ion. Default
is 30 volts.
Isolation Time The dwell time during which the rf field is held
constant in the high mass ejection step. Range
1 - 10 msec.
Excitation Time Time for collision-induced dissociation (CID) by
ion excitation. Range 0 to 1000 msec. The
default is 20 msec.
Modulation Range Mass range over which the rf storage field is
modulated during resonant CID. The range is 0
to 10 steps, default is 2 steps. Value of 0 gives
a fixed rf storage field, i.e., no modulation.
Modulation Rate The number of microseconds that will be spent
at each step. The range is 29 usec to 5600
usec, default is 3000 usec/step.
Number of Frequencies Enter an odd integer in the range of 1 to 121. A
frequency number of one generates a single
supplementary frequency applied to the end
caps. An integer of 3, generates three
frequencies: the resonance frequency and
frequencies 500 Hz above and below that
frequency.
CID Frequency Offset To offset the dissociation frequency. The CID
frequency can be adjusted between 3000 Hz.
Ion Preparation MS/MS
Set up Multiple Reation Monitoring by adding lines to a MS/MS run with different
parent ion masses and different dissociation parameters.
Item Description
Parent Ion Mass (m/z) Integer or exact mass. Parent ion mass range is 50
to 650 m/z.
Isolation Window (m/z) Integral and fractional mass isolation windows are
both accepted. Range is 1.0 to 14.0 m/z and is
mass dependent. The default value is 3.0 m/z.
Waveform Type Nonresonant or resonant excitation.