Default Settings for Predefined AutoSampler Modes
The following table lists the parameter settings used for each of the Predefined
8400/8410 modes using a 10 microliter syringe. If the predefined modes do not
work acceptably for your samples, use the User Defined mode to enter settings
that will work better with your samples. The values listed in this table provide a
starting point for setting the various parameters.
Parameter
Std Split/
Splitless
Std On-
Column
Neat
Volatile
Visco
us
Solvent Plug
Settings:
Solvent plug is not used with predefined modes. If you wish to
use solvent plug injections, you must use User Defined mode.
PreDefined
Settings:
Fill Volume 7.5 uL 7.5 uL Not used Not used Not
used
Fill Strokes 5 5
0
0 0
Sample Air Gap No No No No No
Air Plug after
Sample
1 uL 1 uL 1 uL 1 uL 1 uL
Viscosity
Settings:
Viscosity Delay 0 sec 0 sec 0 sec 6 sec 9.9 sec
Fill Speed 2 uL/sec 2 uL/sec 2 uL/sec 1 uL/sec 1
uL/sec
Inject Speed 50 uL/sec 2 uL/sec 50 uL/sec 1 uL/sec 5
uL/sec
Pre-Inj Delay 0 sec 0 sec 0 sec 0 sec 0 sec
Post-Inj Delay 0 sec 6 sec 0 sec 0 sec 12 sec
Instead of fill strokes Neat mode fills the syringe with sample at 2 uL/sec then
expels it into the waste cup at 50 uL/sec. It does this a total of six times. Then it
fills the syringe with sample at 2 uL/sec and expels it back into the sample vial at
50 uL/sec. This “pumping” action is done three times. After this is completed, the
sample is loaded into the syringe and injected using the parameters in the table.