Item Description
Calibrate Vial 0 Position, Start
Calibration
To start the Vial 0 (or Vial 1) calibration, make
certain a vial with the same form factor and cap
as the ones you will be using for your analyses
is placed in the 0 position of the carrousel (or
the Vial 1 position of the CP-8410). Press the
Start Calibration button.
Calibrating the Vial 0 (or Vial 1) position
involves manipulating the positions of three
different components. The object of the
calibration is to position the syringe exactly over
the center of the sampling target area of the
vial. Upon starting the calibration procedure,
the CP-8400/8410 will rotate the tower, rotate
the carrousel, and lower the syringe sled to the
approximate location of Vial 0 (or Vial 1 if
CP-8410). After the CP-8400/8410 has found
the approximate Vial 0 (or Vial 1) position,
visually check that the syringe sled is not
resting on the vial top. If it is use the Syringe
Up button. Press the button until the syringe
sled is clear of the vial top by approximately 1
mm. Start the calibration process by adjusting
the tower position
Calibrate Vial 0 Position, Save
Calibration
When you have completed the Vial 0 (or Vial 1)
calibration, press the Save Calibration button.
Calibrate Vial 0 Position, Tower
Counterclockwise
Moves the tower 10 steps in the
counterclockwise direction each time the button
is pressed.
NOTE: One basic rule associated with this
calibration concerns rotation of the tower and
carrousel; to assure consistent operation,