Type 11 Detector EFC Status:
Item Description
Type 11
Displays the Detector EFC type. Detector EFC Type 11
has 3 channels and is used for FID detector gas control.
Ready Yes or No.
The Detector EFC is Ready (Yes) if the setpoints have
been achieved and stabilized. The Detector EFC is Not
Ready (No) if readiness to start a run has not been
achieved or if, during a run, the Detector EFC becomes not
ready or goes “out of tolerance”. Some reasons the GC
may go Not Ready during a run are: the temperature
program is too aggressive, the oven door is opened, the
coolant runs out, the power line voltage drops, etc.
Fault Yes or No.
Displays if a fault in the Detector EFC component has
occurred
Ch 1 (Make up) Set,
Actual
Displays, in ml/min, the programmed and actual channel 1
flow rate.
Ch 2 (H2) Set, Actual Displays, in ml/min, the programmed and actual channel 2
flow rate.
Ch 3 (Air) Set, Actual Displays, in ml/min, the programmed and actual channel 3
flow rate.
Type 12 Detector EFC Status:
Item Description
Type 12
Displays the Detector EFC type. Detector EFC Type 12
has 3 channels and is used for TSD detector gas control.
Ready Yes or No.
Fault Yes or No.
Ch 1 (Makeup) Set,
Actual
Displays, in ml/min, the programmed and actual channel 1
flow rate.
Ch 2 (H2) Set, Actual Displays, in ml/min, the programmed and actual channel 2
flow rate.
Ch 2 (Air) Set, Actual Displays, in ml/min, the programmed and actual channel 3
flow rate.