Axis BTA 2100D Manual do Operador Página 164

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 597
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 163
162
Chemical Ionization Using Fixed Parameters
The reaction parameters that are normally set by the Auto Control can also be
set manually. For Chemical Ionization using fixed parameters, the instrument
uses the Ionization Time entered in the method editor. The reaction time that
will be used is determined as follows:
TimeIonizationx
TimeIonizationMaximum
TimeReactionMaximum
=TimeReaction Set
where the maximum reaction and ionization times are entered below the
Ionization Time.
Using Liquid CI Reagents
Introduction
Liquids may be used with the Saturn GC/MS as CI reagents in addition to the
standard CI reagent gases. The liquid is placed in a reservoir and connected in
place of the tank of CI gas. The vapor pressure of the liquid reagent is usually
sufficient to provide a constant, low flow to allow operation in the CI mode.
Hardware
The figure below shows a functional block diagram of the CI plumbing setup for
using liquid reagents.
Liquid
Reservoir
CI Shutoff
Valve
Restrictor
Restrictor
CI Solenoid
To Manifold
CI Needle
Valve
To Roug h ing Pum p
Functional Block Diagram of Liquid CI Plumbing
Vista de página 163
1 2 ... 159 160 161 162 163 164 165 166 167 168 169 ... 596 597

Comentários a estes Manuais

Sem comentários