
374
Item Description
Calibrate Vial 0 Position, Start
Calibration
To start the Vial 0 (or Vial 1) calibration, make
certain a vial with the same form factor and cap
as the ones you will be using for your analyses
is placed in the 0 position of the carrousel (or
the Vial 1 position of the CP-8410). Press the
Start Calibration button.
Calibrating the Vial 0 (or Vial 1) position
involves manipulating the positions of three
different components. The object of the
calibration is to position the syringe exactly
over the center of the sampling target area of
the vial. Upon starting the calibration
procedure, the CP-8400/8410 will rotate the
tower, rotate the carrousel, and lower the
syringe sled to the approximate location of Vial
0 (or Vial 1 if CP-8410). After the
CP-8400/8410 has found the approximate Vial
0 (or Vial 1) position, visually check that the
syringe sled is not resting on the vial top. If it is
use the Syringe Up button. Press the button
until the syringe sled is clear of the vial top by
approximately 1 mm. Start the calibration
process by adjusting the tower position
Calibrate Vial 0 Position, Save
Calibration
When you have completed the Vial 0 (or Vial 1)
calibration, press the Save Calibration button.
Calibrate Vial 0 Position, Use
Mouse Buttons to Rotate Tower
Clicking this button will display the following
dialog box.
Comentários a estes Manuais