309
1177 Injector Status:
Item Description
Ready Yes or No.
The 1177 injector oven is Ready (Yes) if the setpoints have
been achieved and stabilized. The 1177 injector oven is Not
Ready (No) if readiness to start a run has not been achieved or
if, during a run, the oven becomes not ready or goes “out of
tolerance”.
Fault Yes or No.
Displays if a fault in the 1177 injector component has occurred.
1177 Oven On or Off.
Displays if the 1177 injector oven component is turned off or on
at the 3800 GC. The 1177 injector is an isothermal injector.
Setpoint 50 to 450 °C
Displays the 1177 injector oven programmed temperature
setting in the active method.
Actual 50 to 450 °C
Displays the actual temperature of the 1177 injector oven at
the 3800 GC.
Split State On or Off.
Displays the current split state of the 1177 injector method.
1079 Injector Status:
Item Description
Ready Yes or No.
The 1079 injector oven is Ready (Yes) if the setpoints have
been achieved and stabilized. The 1079 injector oven is Not
Ready (No) if readiness to start a run has not been achieved
or if, during a run, the oven becomes not ready or goes “out
of tolerance”. Some reasons the GC may go Not Ready
during a run are: the temperature program is too aggressive,
the oven door is opened, the coolant runs out, the power line
voltage drops, etc.
Fault Yes or No.
Displays if a fault in the 1079 injector component has
occurred.
1079 Oven On or Off.
Displays if the 1079 injector oven component is turned off or
on at the 3800 GC.
Coolant On or Off.
Displays if the 1079 injector oven coolant is enabled in the
active method.
Setpoint -99 to 450 °C
Displays the 1079 injector oven programmed temperature
setting in the active method.
Actual -99 to 450 °C
Comentários a estes Manuais