Item Description
always approach the final calibration position
with clockwise movements. To help facilitate
this rule, the carrousel will backup 10 steps
when a counterclockwise action is selected but
will move one step when a clockwise tower
rotation is selected.
Calibrate Vial 0 Position, Tower
Clockwise
Moves the tower 1 step in the clockwise
direction each time the button is pressed.
NOTE: One basic rule associated with this
calibration concerns rotation of the tower and
carrousel; to assure consistent operation,
always approach the final calibration position
with clockwise movements. To help facilitate
this rule, the carrousel will backup 10 steps
when a counterclockwise action is selected but
will move one step when a clockwise tower
rotation is selected.
Calibrate Vial 0 Position,
Carrousel Counterclockwise
Moves the carrousel 10 steps in the
counterclockwise direction each time the button
is pressed.
NOTE: One basic rule associated with this
calibration concerns rotation of the tower and
carrousel; to assure consistent operation,
always approach the final calibration position
with clockwise movements. To help facilitate
this rule, the carrousel will backup 10 steps
when a counterclockwise action is selected but
will move one step when a clockwise tower
rotation is selected.
Calibrate Vial 0 Position,
Carrousel Clockwise
Moves the carrousel 1 step in the clockwise
direction each time the button is pressed.
NOTE: One basic rule associated with this
calibration concerns rotation of the tower and
carrousel; to assure consistent operation,
always approach the final calibration position
with clockwise movements. To help facilitate
this rule, the carrousel will backup 10 steps
when a counterclockwise action is selected but
will move one step when a clockwise tower
rotation is selected.
Calibrate Vial 0 Position, Syringe
Up
Use the Syringe Up button and adjust the
syringe sled height so that the needle is just
barely above the vial cap (You should be able
to see about a paper’s thickness of space
between the syringe sled needle guide and the
top of the vial cap. The penetration depth into
the vial is measured from the bottom of the
needle guide, so this height will give you
optimum penetration depth.).
Calibrate Vial 0 Position, Syringe
Down
Use the Syringe Down button and adjust the
syringe sled height so that the needle is just
barely above the vial cap (You should be able
to see about a paper’s thickness of space
between the syringe sled needle guide and the
top of the vial cap. The penetration depth into
the vial is measured from the bottom of the