Item Description
Weighted Regression None, 1/n, 1/nX, 1/X2, or 1/nX2
Sets the weighted regression scheme to be used in
determining calibration coefficients.
Replicate Treatment Keep Replicates Separate or Average Calibration
Replicates
Determines whether new replicates are kept separate
or whether they are averaged into an existing historical
value. When keeping replicates separate, there is a
limit of 10 replicates that can be stored. However, you
can indefinitely continue to average new replicates into
the historically maintained value.
Averaging Weight 0-100%
Sets the weight given to new calibration replicates
when averaging into the historical value.
Replicate Tolerance Always add new replicates, Never add new replicates,
Add replicates when within this tolerance (%)
Determines whether a new replicate should be added
to the calibration curve or whether it should be
ignored. Always add new replicates does no
verification of replicates -- they are always added.
Never add new replicates does not add the replicate
and is seldom used. Add replicates when within this
tolerance (%) only adds a replicate to the calibration
curve if within the tolerance specified. This tolerance is
applied the same way as is the Deviation Tolerance for
a Verification run.
Replicate Tolerance % 0 to 500%
Sets the replicate tolerance value.
Replicate Tolerance Out-
of-Tolerance Action
No Action, Increment Error Count, Terminate Sample
List, Halt Automation
Sets which action is taken when a replicate is of
tolerance. Either does nothing (No Action), adds 1 to
the error count (Increment Error Count), stops the run
and goes to the next SampleList (Terminate Sample
List), or halts automation (Halt Automation) each time
a replicate is out of tolerance.
Calibration Range
Tolerance %
0 to 500%
Sets the calibration range tolerance percentage for
which peaks outside the calibration range plus the
tolerance generate calibration range errors. Zero is
always used for the low end of the calibration range
check, even if the valid calibration range does not
extend to the y axis.
Calibration Range Out-of-
Tolerance Action
No Action, Increment Error Count, Terminate Sample
List, Halt Automation
Sets which action is taken when a replicate is of
calibration range. Either does nothing (No Action),
adds 1 to the error count (Increment Error Count),
stops the run and goes to the next SampleList
(Terminate Sample List), or halts automation (Halt
Automation) each time a replicate is out of calibration
range.