306
Column Oven Status:
Item Description
Ready Yes or No.
The column oven is Ready (Yes) if the setpoints have been
achieved and stabilized. The column oven is Not Ready (No) if
readiness to start a run has not been achieved or if, during a
run, the oven becomes not ready or goes “out of tolerance”.
Some reasons the GC may go Not Ready during a run are: the
temperature program is too aggressive, the oven door is
opened, the coolant runs out, the power line voltage drops, etc.
Fault Yes or No.
Displays if a fault in the column oven component has occurred.
Oven Power On or Off.
Displays if the column oven component is turned off or on at
the 3800 GC.
Coolant On or Off.
Displays if the column oven coolant is enabled in the active
method.
Setpoint -99 to 450 °C
Displays the column oven programmed temperature setting in
the active method.
Actual -99 to 450 °C
Displays the actual temperature of the column oven at the 3800
GC.
Stabilization
Time
0.00 to 10.00 min
Shows the 3800 GC column oven stabilization time in minutes.
The stabilization time counts down from the programmed
method value and then displays 0.00 minutes when the
stabilization period is over.
Enable Column
Oven Coolant at
30 °C to 450 °C
Displays the programmed column oven temperature at which
the column oven coolant is enabled.
Timeout 0.01 min to 999.99 min
Displays the programmed coolant time-out value in the active
method for the column oven.
Timed Out Yes or No.
Displays Yes if the coolant time-out period has elapsed for the
column oven.
Comentários a estes Manuais