Prescan Ionization Time
Prescan Ionization Time parameter as used in MS/MS applications is set to 200
sec.
Default Parameters for Reagent Gases used in Chemical Ionization Mode
Reagent Gas Methane Isobutane Ammonia
Maximum Ionization Time
(sec)
2000 2000 2000
Maximum Reaction Time
(msec)
60 60 60
CI Storage Level (m/z) 13 19 13
CI Background Mass (m/z) 45 65 40
Ejection Amplitude (v) 9.0 15 9
Using Non-Standard CI Reagent Gases
Gases other than methane, ammonia, or isobutane can be used successfully
with the Saturn GC/MS. If you are using a non-standard CI reagent gas, you
need to select the Reagent Gas list item labeled "User Defined". Then you will
need to input the name of the gas and the values for the CI parameters; the
parameter values will not be set up automatically.
NOTE: When you install a new supply of a non-standard reagent gas, you must
adjust the reagent gas pressure. After the pressure has been adjusted properly,
you should check the pressure at least once every day if you plan to operate in
CI on a daily basis. For procedures to adjust and check the pressure of a non-
standard reagent gas, please contact a Customer Support Representative.
How to Optimize the CI Background Mass
The proper setting for the CI background mass eliminates reagent ions which are
left over after the reaction period.
With the CI reagent connected and properly adjusted, click the CI Gas control on
the System Control Manual Control page to open the CI gas valve solenoids.
Turn on the trap and observe the spectrum over the scan range specified in the
acquisition method, (don’t use adjust CI Gas). Look for any ions at the low end of
the range. The figure below shows an acetonitrile spectrum with the CI
background mass set too low (=50 m/z). The 54 ion is clearly visible and has
caused a reduced ionization time. The next figure shows a proper background
mass setting for the same conditions (=65 m/z).