57
Turn OFF main power by placing switch at rear of system into OFF (down)
position. Manually vent the system for at least 5 minutes using the lever on front
panel.
NOTE: DO NOT ATTEMPT to VENT SYSTEM by any other method (via transfer
line, foreline vacuum clamps, etc.). Rapid, improper venting of the system will
cause diffusion pump fluid to backstream into the manifold and ion trap.
Maintenance procedures SHOULD NOT BE attempted until the shutdown
program has finished!
Acquisition Mode
Item Description
Start Acquisition button
Click to start the analysis.
Reset button
Will end the acquisition and reset the
conditions to their initial conditions.
End time button
Click to increase or decrease the end time of
the analysis.
Hide Keypad button
Causes the upper portion of the screen to be
removed so that the lower portion can be
expanded in the display.
Display Box Selector
Allows you to choose among chromatogram,
spectra, or both.
Method button
Used to open the Method Editor with the
active method. You can edit the method and
then return to this screen for analysis.