244
Internal Standard Section
Item Description
Use
Yes or No
If Yes, an internal standard addition will be used. When
internal standard addition is used, the internal standard
solution will be drawn up into the syringe from the specified
solvent vial before the sample is drawn up.
Vial I, II, or III
Selects which solvent vial contains the internal standard.
Volume (uL) 0 to 9.0 µL with 10 uL syringe
0 to 4.9 µL with 5 uL syringe
Specifies the amount of internal standard to be drawn into
the syringe before the sample.
Drawup Speed
(uL/sec)
0.1 to 50.0 µL per second with 10 µL syringe
0.1 to 25.0 µL per second with 5 µL syringe
Specifies the speed at which the internal standard will be
drawn up into the syringe.
Pause Time (sec) 0 to 9.9 seconds
Specifies how long the syringe is to remain in the internal
standard vial after drawing up the internal standard.
Air Gap Yes or No
If Yes, 1 µL of room air will be drawn into the syringe to
create an air gap between the internal standard and the
sample.