Indicate whether or not you will use the installed FID Detector by clicking on the
“On” or “Off” radio buttons.
If you indicate that an FID Detector is to be turned on and heated, select the
desired Temperature setting.
Time-program the detector range and autozero in the spreadsheet.
Select the desired FID Time Constant by clicking on the “Slow” or “Fast” radio
buttons.
Select the desired gas flow rates for the detector EFC.
The first row of the spreadsheet contains the Initial Range and Autozero for the
FID Detector. The FID will switch to this setting when the method is activated,
and will be restored to this setting when each chromatographic run is completed.
Rows after the first row of the spreadsheet contain the time-programmed settings
for the Range and Autozero. They will be switched to the value in each row at the
indicated time.
Methanizer
The Methanizer is programmed from the Detector window. Note that the Front
Methanizer is programmed from the Front Detector tab, the Middle Methanizer
from the Middle Detector tab, and the Rear Methanizer from the Rear Detector
tab.