PFPD Detector
If you indicate that a PFPD Detector is installed, an Oven Power switch, an
Electronics switch, Square Root Mode switch, and a Temperature setting will
appear in the top portion of the window, and additional detector adjustments will
appear below the spreadsheet.
If you indicate that an PFPD Detector is to be turned on and heated, select the
desired Temperature setting.
Indicate whether you will use the installed Detector Oven and Electronics by
clicking on the “On” or “Off” radio buttons.
Enable or disable Square Root Mode.
You can time-program the detector range and autozero in the spreadsheet.
Select the desired PFPD Photomultiplier Voltage, Gate Delay, Gate Width, and
Trigger Level from the available choices.
Select the desired gas flow rates.
The first row of the spreadsheet contains the Initial Range and Autozero for the
PFPD Detector. The PFPD will switch to this setting when the method is
activated, and will be restored to this setting when each chromatographic run is
completed.
Rows after the first row of the spreadsheet contain the time-programmed settings
for the Range and Autozero. They will be switched to the value in each row at the
indicated time.