when it is activated. It also allows the configuration of the 3900 GC to be
changed from the Saturn GC/MS Workstation.
Item Description
Column, Rapid MS Mode On/Off
When a Rapid-MS column is used in the 3900
GC, the Rapid-MS Mode should be set to On.
This will configure the GC to set the proper
parameters for the Rapid-MS column. The
column length and inside diameter entries will
be grayed out.
Column, Length The length of the column in meters. Values
range from 0 to 250 meters.
Column, Inside Diameter Inside diameter of the column in micrometers.
Values range from 0 to 999 micrometers.
Temperature Limit, Column Oven Upper temperature limit of the column oven.
Values range from 50 to 450°C.
Temperature Limit, Injector Upper temperature limit of the injector. Values
range from 50 to 450°C.
Ready-In Sync Polarity Is Ready-In a closed contact (Checked/Un-
checked). Allows you to specify what switch
state corresponds to External Device Ready.
Injector EFC, Splitless Vent Flow The flow the system will vent when the split
valve is left closed for the duration of the run
such as during an on-column injection. This
flow is diverted before the injector and thus
does not influence the splitless nature of the
run. This entry is ignored during split and