Item Description
Volume 0.1 to 10 ul with 10 ul syringe, or
1 to 100 ul with 100 ul syringe
Sets the volume in microliters to be drawn
into the syringe and add to the vial to be
sampled.
Mixes 0 to 100
Sets the number of mixing strokes to be done.
Mix Stroke 0.1 to 10 ul with 10 ul syringe, or
1 to 100 ul with 100 ul syringe
Sets the volume to be drawn into the syringe
in microliters for each mixing stroke.
Washes 0 to 9
Sets the number of times the syringe will be
washed after this step.
Reaction Time 0.0 to 99.9 minutes
Pause time to allow reactions to finish before
the next step of the automix routine or sample
injection is performed.
Add button Adds a new automix line to the end of the
automix routine
Insert button Inserts a new automix line before the selected
line.
Delete button Deletes the selected automix routine line.
Automix First Injection Only When checked, this automix routine will only
be performed before the first injection from
the vial. When unchecked, this automix
routine will be performed before each
injection.
OK button Saves the edits and closes the dialog box.
Cancel Discards edits and closes the dialog box.
Set as Default Automix Saves the highlighted line as the default
entry. New lines will be created with these
values loaded.