Ionization Mode - EI Fixed
Item Description
Mass Range Segment
Breaks
The EI scan is made up of between 1 and 6 scan
segments, each holding a contiguous mass-range
segment. Overall, the set covers the entire mass range
of 10 to 650 m/z. Only the intermediate segment
boundaries can be changed by the user.
Ionization Storage
Level
The rf voltage used during the ionization period. Values
from 5 to 150 m/z.
Ionization Time
Factor
The Ionization Time Factor, given as a percent, is a
number that is multiplied by the calculated ionization
time (determined by the AGC pre-scan pulse) to give the
actual ionization time. Values from 1 to 999%.
Ionization Time
Sets the fixed ionization time. Values from 10 to 65000
sec.
Max Ionization Time
Sets the maximum ionization time used by the system.