Axis BTA 2100D Manual do Operador Página 432

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 597
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 431
430
Specify the calibration type.
Up to 10 separate replicates.
Percent of new versus historical data.
Refer to the on-line help in Interactive Graphics for a description of all fields in
this window.
Refer to the Regulatory Compliance Manual for a description of weighting
options.
Peak Table
The Peak Table contains amounts for each level (the number of which is
specified in the Calibration Setup), and the internal standard peak designations.
Specify internal standard peaks.
Refer to the appropriate standard peak.
Enter amounts for each level.
Automated Calibration
You calibrate your Method by injecting calibration samples using System Control.
Calibration samples are specified in the SampleList.
To clear previously stored calibration data, enter a New Calib Block line.
Select the Calibration sample type.
Enter the calibration level for this injection.
You may inject more than one replicate for each level.
Vista de página 431
1 2 ... 427 428 429 430 431 432 433 434 435 436 437 ... 596 597

Comentários a estes Manuais

Sem comentários