Two filaments are provided to reduce down time for maintenance. The selection
of the filament is made by choosing the radio button. The Axial Modulation
voltage can be changed by using the up/down arrows and then clicking the apply
button.
The Axial Modulation Voltage is applied to the endcap electrodes as described in
the section Saturn Method Editor. Proper setting of this voltage is critical to
resolution and sensitivity in the ion trap. In most cases, the best Axial Modulation
voltage is the lowest value at which there is still good resolution of the 131/132
and 414/415 cal gas peaks. For most instruments this value will be between 2
and 4 volts.
The Axial Modulation voltage must be adjusted to its proper value before the
AGC target TIC is determined and before the mass calibration is done.
If the value of the Axial Modulation voltage is too low, high molecular weight ions
will not be observed. If the value for the Axial Modulation voltage is too high, the
peak widths for low molecular weight ions will be broadened and mass
misassignments may occur.
NOTE: The mass calibration must be done if the axial modulation setting is
changed. See the autotune section for mass calibration.
Adjustments
The third tab is the Adjustments tab. Clicking on this tab will bring up the screen
below.
This tab is used to adjust the Cal gas pressure, the CI reagent gas pressure,
select the CI reagent, and to adjust the rf ramp.
When you select Adjust rf Tuning Ramp, you obtain a screen from which you
can determine if the rf voltage is in tune throughout the entire mass range of the
Saturn GC/MS. This is done by observing the rf modulator response.