320
Type 13 Detector EFC Status:
Item Description
Type 13
Displays the Detector EFC type. Detector EFC Type 13 has
1 channel and is used for either ECD or TCD detector gas
control.
Ready Yes or No.
Fault Yes or No.
Ch 1 (Makeup) Set,
Actual
Displays, in ml/min, the programmed and actual channel 1
flow rate for TCD detector gas control.
Ch 1 (Reference) Set,
Actual
Displays, in ml/min, the programmed and actual channel 1
flow rate for ECD detector gas control.
Type 14 Detector EFC Status:
Item Description
Type 14
Displays the Detector EFC type. Detector EFC Type 14
has 2 channels and is used for TCD detector gas control.
Ready Yes or No.
Fault Yes or No.
Ch 1 (Make up) Set,
Actual
Displays, in ml/min, the programmed and actual channel 1
flow rate.
Ch 2 (Reference) Set,
Actual
Displays, in ml/min, the programmed and actual channel 2
flow rate if the channel 2 gas type is a Helium, Nitrogen, or
Argon reference.
Ch 2 (Make up) Set,
Actual
Displays, in ml/min, the programmed and actual channel 2
flow rate if the channel 2 gas type is a Helium, Nitrogen, or
Argon make-up.