The informational fields displayed in the above dialog are taken from the
currently executing method.
The Scan Mode
describes the ionization mode of the current segment, for
example, EI-Auto.
The Ion Prep.
is the ion preparation method of the current segment, such as
MS/MS.
The Range
will show the mass range being scanned by the current segment.
Operating Conditions
The Operating Conditions located to the far left of the Saturn Module window
allows you to observe the current instrument information.
The Mode State
will indicate what function the Saturn MS is currently performing
such as scanning or adjusting cal gas.
The Fault State
will indicate if a fault has occurred.
The Ion Time
indicates the ionization time of the last scan.
The Ion Count
indicates the ion count of the last scan. If profile mode is selected,
this is the total ion count. If centroid mode is selected, this is the reconstructed
ion count.
SetPoints
The SetPoints tab allows us to select the filament and the axial modulation
voltage. Clicking on that tab will bring up the following screen.