Item Description
needle guide, so this height will give you
optimum penetration depth.).
Calibrate Injector Position, Start
Calibration
Press the Start Calibration. The CP-8400/8410
will rotate the tower so that it is over the Injector
position and lower the syringe sled so that it is
over the injector. If the needle guide descends
so that it is either resting on the inject switch or
below the needle cone on the inject switch, use
the Syringe Up button and move the sled up
until the needle guide is just above the injector
cone.
Calibrate Injector Position, Save
Calibration
Press the Save Calibration softkey when the
Injector calibration is complete.
Calibrate Injector Position, Tower
Counterclockwise
Using the Tower Counterclockwise button
rotates the tower counterclockwise 10 steps.
Position the tower directly over the injector
Inject Switch Locator cone.
Calibrate Injector Position, Tower
Clockwise
Using the Tower Clockwise button rotates the
tower clockwise 1 step. Position the tower
directly over the injector Inject Switch Locator
cone.
Calibrate Injector Position,
Syringe Up
Using the Syringe Up button adjust the height
of the needle guide such that the sled is just
resting on or very slightly depressing the
Injector Switch and is covering the locator cone.
Calibrate Injector Position,
Syringe Down
Using the Syringe Down button adjust the
height of the needle guide such that the sled is
just resting on or very slightly depressing the
Injector Switch and is covering the locator cone.
Change Syringe, Start Syringe
Change
Selecting Start Syringe Change will cause the
tower to rotate such that the user will have easy
access to the syringe for routine service and
syringe removal and replacement. The buzzer
will sound before any tower movement occurs.
Press the Finish Syringe Change softkey upon
completion of the syringe maintenance.
Change Syringe, Finish Syringe
Change
Selecting Finish Syringe Change will cause the
tower to return to its previous position and
state. The buzzer will sound prior to movement
as designated by the “Buzzer at start of cycle”
setting.
Change Syringe, Clear Plunger
Counter
Selecting this softkey causes the Plunger
stroke count to be reset to 0.
Use Mouse Buttons to adjust
positions of Tower, Syringe,
and/or Carrousel
Checking this box, changes the Calibrate
AutoSampler Dialog Box for the positions to be
adjusted with the mouse.
Close Closes the dialog box.