
195
Mass Range Segment Breaks
Each EI scan is divided by default into four mass segments: 10 to 99 m/z, 100 to
249 m/z, 250 to 399 m/z and 400 to 650 m/z. For certain applications, the mass
ranges in these segments can be adjusted. Additional segments (maximum of 6
segments) can be added by clicking on the Insert button.
Ionization Storage Level
Each segment has its own ionization storage level. This corresponds to an rf
voltage that is used to hold ions in the trap during the ionization period. It affects
ion storage in two ways— the storage efficiency of higher mass ions increases
as the level increases, while lower mass ions are not stored if their mass falls
below the cutoff. With AGC on, the default storage level is set to 35 m/z, causing
all ions above 35 m/z to be stored. This value gives good storage efficiency for
ions up to 650 m/z, while not storing the 18 and 19 m/z ions resulting from water
vapor.
For certain analyses, the storage voltage may be increased to prevent the
storage of ions produced from the solvent. For example, to exclude ions
produced from methanol (mass 32 and below), the storage voltage may be set to
38 m/z for all segments. Keep the EI Background mass greater than or equal to
the segment ionization storage voltage.
Ionization Time Factor
The Ionization Time Factor is a secondary factor that determines how well the
Saturn GC/MS is tuned. The Ionization Time Factor, given as a percent, is a
number that is multiplied by the calculated ionization time (determined by the
AGC pre-scan pulse) to give the actual ionization time. The default value of the
time factor is 100%. Ionization Time Factor determines the actual ionization time
for each segment; thus, it determines the number of ions in the ion trap for that
segment
The calculated ionization time, determined by the AGC pre-scan pulse, is
generally the optimum value for segment 2. At lower masses (segment 1),
however, an actual ionization time of as low as 75% (and never lower than 50%)
of the calculated ionization time will sometimes provide a better tune. And for
higher masses (segments 3 and 4), an actual ionization time of up to 125% (and
never higher than 200%) of the calculated value sometimes provides better
results.
As the ionization time factor is increased (actual ionization time increased), the
peak height increases. However, the resolution of adjacent mass peaks may
decrease and the peak shape might distort.
Target TIC Value
In EI mode with AGC on, the target TIC value determines how many ions AGC
allows into the ion trap during the calculated ionization time. While increasing the
target TIC value will increase peak heights, increasing it too far will result in a
loss of mass resolution. This would be observed as mass misassignments and/or
incorrect isotope abundance ratios. For full-scan EI operation, a target TIC of
20,000 is a good starting point. Under special circumstances, e.g., MS/MS, a
much lower target TIC is required. As the m/z distribution of the stored ions
becomes smaller, fewer ions can be trapped without a loss of resolution.
Comentários a estes Manuais