TSD Detector
If you indicate that a TSD Detector is installed, an Oven Power switch, an
Electronics switch, a Temperature setting, a Bead Current setting and a Time
Constant setting will appear.
If you indicate that an TSD Detector is to be turned on and heated, select the
desired Temperature and Bead Current setting.
Indicate whether or not you will use the installed TSD Detector by clicking on the
“On” or “Off” radio buttons.
You can time-program the detector range, autozero, and bead power in the
spreadsheet.
Select the desired TSD Time Constant by clicking on the “Slow” or “Fast” radio
buttons.
Select the desired gas flow rates for the detector EFC.
The first row of the spreadsheet contains the Initial Range, Autozero, and Bead
Power settings for the TSD Detector. The TSD will switch to this setting when the
method is activated, and will be restored to this setting when each
chromatographic run is completed.
Rows after the first row of the spreadsheet contain the time-programmed settings
for the Range, Autozero, and Bead Power. They will be switched to the value in
each row at the indicated time.