243
Item Description
Vial I, II, or III
Specifies which of the three solvent vials will be used by the
Default Cleaning
Volume (uL) 0 to 10.0 microliters with 10 uL syringe
0 to 5.0 microliters with 5 uL syringe
Specifies the amount of cleaning solvent that will be drawn up
with each syringe cleaning stroke.
Strokes 0 to 10
Specifies the number of times the cleaning solvent will be drawn
up into the syringe and expelled into the waste cup.
Speed (uL/sec) 0.1 to 50.0 µL per second with 10 µL syringe
0.1 to 25.0 µL per second with 5 µL syringe
Specifies the speed at which the cleaning solvent will be drawn
up into the syringe and the speed at which it is expelled into the
waste cup.
Clean Mode Section
Item Description
Pre-Inj Solvent
Flushes
0 to 99
Specifies the number of times each selected cleaning
solvent will be drawn up into the syringe and expelled into
the waste cup before flushing with next cleaning solvent (if
more than one cleaning solvent source is specified) or the
sample.
Pre-Inj Sample
Flushes
0 to 99
Specifies the number of times the sample will be drawn up
into the syringe and expelled into the waste cup before the
syringe is loaded for injection.
Post-Inj Solvent
Flushes
0 to 99
Specifies the number of times each selected cleaning
solvent will be drawn up into the syringe and expelled into
the waste cup before flushing with next cleaning solvent (if
more than one cleaning solvent source is specified).
Clean Solvent Source I, II, III, I & II, I & III, II & III, or I & II & III
Selects which solvent vial or sequence of solvent vials will
be used to flush the syringe before and after injection.
Comentários a estes Manuais