Axis BTA 2100D Manual do Operador Página 340

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 597
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 339
338
Item Description
Vial I, II, or III
Selects which solvent vial to use for the solvent plug.
Volume (uL) 0 to 10.0 microliters with 10 uL syringe
0 to 5.0 microliters with 5 uL syringe
Specifies the amount of solvent to be drawn into the
syringe before the sample or internal standard.
Drawup Speed (uL/sec) 0.1 to 50.0 microliters per second with 10 uL syringe
0.1 to 25.0 microliters per second with 5 uL syringe
Specifies the speed at which the solvent will be
drawn up into the syringe.
Pause Time (sec) 0 to 9.9 seconds
Specifies how long the syringe is to remain in the
solvent vial after drawing up the solvent.
Air Gap Yes or No
If Yes, 1 microliter of room air will be drawn into the
syringe to create an air gap before the solvent plug.
User Defined Settings
Vista de página 339
1 2 ... 335 336 337 338 339 340 341 342 343 344 345 ... 596 597

Comentários a estes Manuais

Sem comentários