266
3800 GC Data Acquisition
Item Description
Detector Bunch Rate Powers of 2 from 1 to 128 points.
Sets the detector bunch rate in numbers of points;
shows the resulting frequency in Hz. With increased
bunch rates, less disk space is required for storing the
data. With decreased rates, more data is collected.
Bunch rate should be set appropriately for the
narrowest peak of interest in your chromatogram.
Monitor Length 16 to 128 bunched points.
Sets the number of points collected during the
monitoring period, shows the resulting time in seconds.
Used to set the number of data points used to calculate
the baseline noise. Generally, a larger value entered
here will yield a lower noise calculation.
FID/TSD Detector Full
Scale
(Font, Middle, Rear)
1V, 10V, 100V, 1000V.
Specifies the maximum detector value that will be
processed by the Varian MS Workstation. The dynamic
range of the FID and TSD detectors are equivalent to
1000V full scale, but if your peaks of interest fall
significantly below this value, you may wish to limit the
full scale value here. Since the attenuation setting used
by the Standard Report application to scale the
chromatogram printout is based on the detector's full
scale value, a lower full scale value will allow you to
scale smaller peaks with greater resolution.