Tips for Using SIS
Observe these recommendations for best results when using the SIS application.
When selectively storing ions that are easily fragmented by collision-induced
dissociation (CID), use wide mass isolation ranges and low amplitude
waveforms.
As the frequency notches become increasingly narrow and the voltage of the
frequency components increases, the ions that you wish to store acquire
sufficient energy to dissociate or be ejected. As a result, a small fraction of
these ions may be lost due to dissociation or ejection.
Target TIC values less than 10,000 are recommended when using SIS
applications, especially with a narrow mass window. Values in excess of
10,000 may cause space charge build-up, which can decrease the linear
range for quantitation and result in loss of resolution.
The MS “q” Calculator
For convenience, The MS q Calculator is available in all Ion Preparation Options.
Storage rf values are often reported in terms of the Mathieu “q” parameter in the
scientific literature. The value of the “q” parameter determines the stability of the
parent ion trajectory. More stable trajectories allow higher excitation voltages to
be applied before ions are ejected from the trap. It has been empirically observed
that a “q” value of 0.4 provides an optimum yield of product ions. This calculator
utility determines the corresponding CID storage rf value (m/z) following user
entry of an ion mass (m/z) and desired “q”.