Axis BTA 2100D Manual do Operador Página 193

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 597
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 192
191
The ion current measured during the AGC prescan is known as the AGC peak
area. The AGC peak area is proportional to the number of ions formed during the
prescan ionization. The AGC software uses the measured ion current as a
reference to optimize the ionization time for each scan as a function of analyte
concentration.
Scan Segment Setup Period
The prescan is followed by the first segment of the six scan segments. Each
segment can be divided into two parts: a setup period and a scanning period.
Each setup period can be subdivided into the following four stages:
First stage: The rf voltage is set to near 0V for 2 milliseconds. This setting
allows the rf voltage generator circuit to stabilize while clearing the trap cavity of
all ions.
Second stage: The rf voltage is set to, and stabilizes at, the storage voltage,
selected to store, or trap, all ions of interest in the ion trap cavity.
Third stage: First, the electron gate is turned ON. Energetic electrons from the
heated filament are accelerated into the ion trap, where they interact with sample
molecules and sample ions are formed. After the ionization period, the electron
gate is turned OFF, and additional electrons are prevented from entering the ion
trap.
The newly formed sample ions are then permitted to stabilize in their orbits at the
storage voltage. This stabilization period is referred to as the cool time. The
length of this period is fixed.
Fourth stage: The electron multiplier is made ready, and the rf voltage is
ramped quickly from the storage voltage to a value appropriate to begin the mass
scanning of the segment.
Mass Scanning
During mass analysis, the rf voltage is ramped from a low voltage to a higher
voltage.
The ion trap operates in the resonance ejection mode. In this mode, at low rf
voltage (for example, at the storage voltage) most ions have stable oscillations in
the ion trap cavity. However, as the rf voltage is increased, ions of progressively
greater mass-to-charge ratios become unstable and are ejected from the cavity.
The voltage at which an ion is ejected from the ion trap cavity is defined as its
resonance voltage. At a voltage below its resonance voltage, an ion travels in a
stable orbit, oscillating along all three axes. At or above its resonance voltage,
the trajectory of the ion becomes unstable. As the rf voltage increases above the
resonance voltage, the amplitude of the oscillation of the ion increases most
rapidly in the axial direction. As a result, when ions are no longer stable in the ion
trap cavity, they are ejected from the cavity in a tightly collimated beam upward
and downward. The ejection of ions of each mass-to-charge ratio occurs over a
very short time period. One half of the ions go up; these ions strike the top
endcap and are neutralized. The other half of the ions go down and pass through
the holes in the bottom endcap; many of these ions strike the electron multiplier
and are detected.
The rf voltage is ramped at a constant rate of approximately 5600 m/z per sec.
One complete scan at this fixed rate produces a microscan. Depending upon the
Vista de página 192
1 2 ... 188 189 190 191 192 193 194 195 196 197 198 ... 596 597

Comentários a estes Manuais

Sem comentários