Click Show Keypad to display the keypad.
Status and Control Window
You are now ready to program the GC/MS to perform fully automated data
acquisition. Initially the MS Module Window will look similar to the figure below. If
you do not see this screen, go to the pull down menu labeled Instrument, and
select the item “Saturn GC/MS”.
The upper left side of the window shows the ion trap schematic, which allows the
ion trap to be turned on. If the trap is displayed in red, it is currently disabled. The
trap is disabled if the current segment is a filament/multiplier delay segment, one
with an ionization mode of none, or if a fault has occurred. If the ion trap icon is
not currently disabled, clicking in the center of the ion trap will turn on the trap.
The black schematic will turn green to denote that it is on, as displayed below.