Calibration Verification
You may wish to periodically check the validity of your calibration data as you
perform analytical injections. You do so by injecting a verification sample with
known amounts for each compound and comparing them to amounts calculated
from the calibration curves. If the amounts deviate more than a given tolerance, a
verification failure occurs and a number of failure actions can take place.
Verification failure options are specified in the Verification Setup window in the
Data Handling Method section.
Specify the tolerance.
Select the failure action. Verification failures are documented in the results
report.
Verification samples are entered into the SampleList (shown) or RecalcList.
Select the Verification sample type.
Specify the calibration level corresponding to the known amounts in the sample.
Refer to the on-line help in Report for details on the format of the results report
for Verification runs.