572
File Revision Settings
Item Description
Require Revision Log entry when
changes to Methods are saved.
When checked, the user is prompted
for a description of changes when a
Method file is altered and saved. This
Revision Log appears in the Notes
portion of the File Open dialog box, is
listed in the Method Builder application
window when the Method is open, and
is included in the Method printout.
Update Revision Log in Data Files with
recalculation date, time and Method
name.
When checked, Data Files are
updated with a time stamp and Method
name when they are recalculated
(either from System Control or from
Interactive Graphics). The Data File
Revision Log can be included in
printed reports.
Enable results deletion from "Open Data
File" dialog boxes.
When checked, a button appears in
the Open Data File dialog box allowing
results to be deleted from a specified
channel of a Data File. Results
deletion is logged in the Data File's
Revision Log. This option only affects
standard GC results. GC/MS results
will not be deleted.