Initial Tangent Height % 0 to 100
Sets the percentage of a mother peak's size below
which a given peak is considered a tangent peak. Used
to assign tangent peak status as a percentage of the
mother peak size. Can be used to force a peak to be
integrated as a tangent or as a fused peak. This value is
overridden if you time program a Tangent Height
Percentage event.
Monitor Noise:
Item Description
Monitor Noise Before Every Run, Once at Start of Method, Fixed Value
Sets times at which the Workstation monitors noise. The
workstation can monitor noise before each run, once at the
start of each Method, or use a user-specified value between
1 and 10,000 (units depend on detector type).
Peak Measurement:
Item Description
Measurement Type Area, Height, Square Root of Height
Sets the method used to measure the peaks.
Initial Peak Reject Value 0 to 1,000,000
Sets the minimum size peak to be identified. Used
to eliminate very small peaks from the report, to
ensure that insignificant peaks are not mistaken for
peaks of interest. This value is overridden if you
time program a Peak Reject value.
Peak Result Calculation:
Item Description
Report Unidentified Peaks checked or not checked
If checked, tells the Workstation to report all peaks,
whether identified or not.
Report Missing Peaks checked or not checked
If checked, tells the Workstation to report any peaks
in the peak table that were not found. Used to see
which peaks should have been found in the sample
but were not.
Normalize Results checked or not checked
If checked, results for each analyte will be expressed
as a percentage amount relative to the total amount
of all components of the sample. This checkbox is
enabled only for external standard and internal
standard calculations.