275
Indicate whether or not you will use the installed Valve Oven by clicking on the “On”
or “Off” radio buttons.
Use the spreadsheet to build a time program to turn the programmable valves on and
off.
Specify the isothermal temperature of each Valve Oven.
The first two rows of the spreadsheet are added automatically. Use the first row
to indicate how each Valve is used, by clicking on the arrow in the top cell of
each Valve column. Select from the choices displayed in the combo box.
Configure each Valve to match the setup on the 3800 GC that will be running the
method.
The second row of the spreadsheet contains the Initial setting for each of the
seven Valves. Each Valve will be switched to this setting when the method is
activated, and restored to this setting when each chromatographic run is
completed.
Rows after the second row of the spreadsheet contain the time-programmed
settings for each of the seven Valves. Each Valve will be switched to the setting
in each row at the indicated time.