Detector Window
The Detector window contains up to three Detector Programs (Front, Middle,
Rear). Use the Detector Programs to specify the operating temperatures, gas
flow rates and parameters of the detectors.
Use the tabs to select which Detector Program you wish to edit.
Select the type of Detector (FID, TSD, TCD, ECD, or PFPD) at each position, or
select “None” if no Detector is installed at that position.
Detector EFC modules
The detector Electronic Flow Control modules (EFC), if installed on your 3800
GC, are detector-specific and can be programmed from the Detector window.
Select the type of Detector EFC at each position, or select “None” if no Detector
EFC is installed at that position. Use the detector EFC program in each detector
window (Front, Middle, Rear) to specify the gas flow rates for each module.
Note that selecting the make-up gas and the auto-calibration of the module can
be performed from the Detector EFC Setup and configuration screens on the
3800 GC.
FID Detector
If you indicate that an FID Detector is installed, an Oven Power switch, an
Electronics switch, a Temperature setting and a Time Constant setting will
appear.