Axis BTA 2100D Manual do Operador Página 441

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 597
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 440
439
Identifying Peaks
Peaks are identified with the use of the Peak Table in the Data Handling section
of the Method. The main purpose of the Peak Table is to list the names of your
compounds and their expected retention times. Since retention times may vary
slightly from run to run, you can specify a window of time during which the
Workstation will try to identify a particular peak. The peak window is the actual
span of time on the chromatogram that the software searches for the peak,
centered around the retention time you entered for that peak. If the peak falls
within that time window, it will be identified, otherwise it will not be identified. You
may choose to report unidentified peaks, but these peaks cannot be accurately
quantified (refer to the on-line help in Interactive Graphics for a description of the
unidentified peak factor).
If your peaks tend to drift, you can automatically modify their peak windows by
marking a peak as a reference peak. Non-reference peaks will have their peak
windows adjusted by the same percentage that the reference peaks have
deviated from their expected retention times, accounting for the direction of
change.
Interactive Graphics provides an easy way to build a Peak Table using your pilot
run. You select the Edit MethodFill Peak Table menu item, and one by one
click on the peaks in your pilot run which you want to identify in your production
runs. As you click on a peak, a new entry is added to the Peak Table with the
retention time and a default name. When you have added all the peaks of
interest, you can change the default names to real compound names.
Building a Calibration Curve
In order to determine absolute amounts of an analyte, you need to compare the
peak area from a run in which the amount is unknown to a run in which you have
injected a known amount of the compound. A sample that contains known
amounts of the compounds to be analyzed (analytes) is called a calibration
standard. When you inject a calibration standard, you designate the run as a
Calibration Run. You enter amounts (for each analyte in the calibration standard)
into the Peak Table. Note that this is the same table used for identifying your
analytes.
Sometimes you will want to do several calibration runs, and in each successive
calibration run you inject different calibration standards with different amounts of
each analyte. This is called multipoint or multilevel calibration. The Varian MS
Workstation allows you to calibrate with up to ten levels. Each level represents
the amount of each analyte injected for each of the different calibration
standards. Multilevel calibration can be more accurate than single level
calibration, because you calibrate over large range of analyte amounts. For more
accuracy, you may want to repeat injections at a given level. This is a way of
averaging out variations from injection to injection.
Regardless of the number of calibration runs you perform, the result of the
calibration runs is a calibration curve. The calibration curve for an analyte is
generated by plotting the peak size on the y axis and the injected amount of the
analyte on the x axis for each calibration run you have performed, and then
calculating the best line through the points. There are options to do a linear,
quadratic or cubic fit through the points.
Vista de página 440
1 2 ... 436 437 438 439 440 441 442 443 444 445 446 ... 596 597

Comentários a estes Manuais

Sem comentários